Doi:10.1016/j.radmeas.2007.10.015
Radiation Measurements 43 (2008) 315 – 318
Characteristics of LiF:Mg,Cu,P thermoluminescence at ultra-high dose range
P. Bilskia,∗, B. Obryka, P. Olkoa, E. Mandowskab, A. Mandowskib, J.L. Kimc
a Institute of Nuclear Physics (IFJ), Krakow, Poland
bInstitute of Physcis, Jan Dlugosz University (AJD), Czestochowa, Poland
cKorean Atomic Energy Research Institute (KAERI), Dejoan, Republic of Korea
The behaviour of LiF:Mg,Cu,P (MCP-N) detectors at ultra-high doses up to 500 kGy has been investigated. Some very significant changes of
the glow-curve shape have been revealed. The most important finding is the appearance of a new peak at exposures above 50 kGy, the positionof which shifts toward higher temperatures with increasing dose (from 405 to 470 ◦C at 500 kGy), in contradiction to standard TL models.
The E and s trapping parameters also increase with increasing dose. This new peak appears to have potential application for ultra-high-dosemeasurements. The TL emission spectrum also changes at such high doses with the appearance of several new bands of wavelengths up to800 nm.
2007 Elsevier Ltd. All rights reserved.
Keywords: Thermoluminescence; High dose; Lithium fluoride
IFJ, Krakow. The standard 240 ◦C/10 min annealing cycle wasapplied. Gamma-ray exposures were performed using the high-
LiF:Mg,Cu,P thermoluminescent detectors are now becom-
activity Co-60 source at KAERI and the therapeutic Co-60
ing standard in modern TL dosimetry. Their main virtue, which
source in Krakow. Doses ranged between 0.5 and 500 kGy. At
gained them their present popularity, is their very high sensi-
each dose four to six detectors were used. Thermolumines-
tivity. LiF:Mg,Cu,P detectors are able to measure doses at mi-
cence was measured up to 600 ◦C using a RA'94 reader at the
crogray levels or even below. Another interesting feature is their
IFJ at a linear heating rate of 2 ◦C/s. The reader was equipped
dose response, which does not show supralinearity up to satura-
with a BG-12 filter. Glow-curves were deconvoluted into sin-
tion at about 1 kGy. Very remarkable is also that the glow-curve
gle first-order peaks using GlowFit, a new program recently
shape remains nearly unchanged over several decades: between
developed in Krakow Spectrally
micrograys and a kilogray These properties are
resolved measurements were performed at AJD Czestochowa
quite unlike those of LiF:Mg,Ti.
using a spectrometer connected with a liquid nitrogen-cooled
Most, if not all, earlier studies of this material at high doses
CCD 1024E camera. TL spectra were recorded from room tem-
perature up to 350 ◦C in a vacuum.
were limited to the above-mentioned dose range.
The present paper reports on investigations of properties of
3. Results and discussion
LiF:Mg,Cu,P at doses much higher than 1 kGy, i.e. above thenominal saturation level of this material.
presents the measured glow-curves with peak heights
normalized to unity over the dose range 0.5–5 kGy. It is ap-
2. Material and methods
parent that increasing the dose results in growth of the high-temperature peaks. At 1 and 2 kGy this is observed only as a
All measurements were performed using LiF:Mg,Cu,P
more pronounced tail of the main peak 4, but, at 3 and 5 kGy,
(MCP-N) thermoluminescent detectors manufactured at the
other peaks arise. The glow-curve over this dose range can be
easily and consistently deconvoluted into six first-order peaks
illustrates changes of the glow-curve shape
E-mail address: (P. Bilski).
1350-4487/$ - see front matter 2007 Elsevier Ltd. All rights reserved.
doi:10.1016/j.radmeas.2007.10.015
P. Bilski et al. / Radiation Measurements 43 (2008) 315 – 318
Fig. 1. LiF:Mg,Cu,P glow-curves over the dose range 0.5–5 kGy (peak heights
Fig. 3. LiF:Mg,Cu,P glow-curves over the dose range 5–50 kGy.
normalized to unity). The arrow indicates increasing doses.
Fig. 4. LiF:Mg,Cu,P glow-curves over the dose range 50–500 kGy.
Fig. 2. The deconvoluted LiF:Mg,Cu,P glow-curve after 5 kGy exposure.
observed at the doses ranging from 5 to 50 kGy. Over this dose
peak should not depend on dose. For second-order kinetics
range the previous main peak 4 becomes strongly suppressed
and even in the more general framework of the simple trap
and at 50 kGy it is practically absent. The high-temperature
model, the peak should shift with increasing dose, but towards
peaks still increase with dose and dominate the whole glow-
lower temperatures The new peak may be well
curve. At 50 Gy a new peak, located at temperatures above
fitted with a first-order kinetic function, this deconvolution
400 ◦C, appears. The glow-curves for this dose region cannot
leading to very high values of energy and of the frequency
be successfully deconvoluted even with 10 peaks.
factor (E = 4.47 eV and s = 5.6 × 1028 s−1 at 500 kGy).
Some very interesting features are observed at doses ex-
The values of these kinetic parameters also increase with in-
ceeding 50 kGy The new peak, which appeared at ca.
creasing dose The obtained results, i.e. unphysical
400 ◦C, still grows with increasing dose but, surprisingly, its
values of frequency factors, as well as the anomalous shift
position (temperature) also becomes dose dependent—shifting
of the high-temperature TL peak with dose, clearly indicate
towards higher temperatures with dose. The difference in the
that the classical first-order kinetic theory does not apply in
peak position between 500 and 50 kGy is as much as 60 ◦C. It
this case. Therefore, the obtained values presumably have
should be emphasized that such an effect is not predicted by
no physical meaning. Thus, a question arises concerning a
the standard TL models. The position of a first-order kinetic
possible theoretical account for this phenomenon. No one
P. Bilski et al. / Radiation Measurements 43 (2008) 315 – 318
Fig. 5. Position (open circles, left-hand scale) and kinetic energy (full circles,right-hand scale) of the new high-temperature peak in LiF:Mg,Cu,P vs. dose.
Fig. 6. Emission spectra of LiF:Mg,Cu,P measured at doses 0.0065, 20 and500 kGy, integrated over the temperature range (60–350) ◦C.
justification, based on standard kinetic models, is known to theauthors.
However, an explanation of this unusual behaviour of the
new peak is proposed by in an accompa-nying paper, whereby traps and recombination centres may beconsidered as a system of interacting clusters. With increasingdose the size of the clusters grows, simultaneously increasingthe effective activation energy and shifting the TL peak towardshigher temperatures. The very-high-frequency factors are pro-duced by a cascade detrapping mechanism due to simultaneousoccurrence of localized and delocalized transitions. One shouldnote that the nature of these ‘defect clusters' is unknown at thismoment, and the problem should be studied in the future.
Examination of emission spectra also revealed interesting
features. Within the spectrum, which for lower doses consists ofa narrow peak at about 350–380 nm, several longer-wavelengthbands gradually appear with increasing dose. At 20 kGythey are observed as a new emission peak at about 550 nmAt the highest exposure, the emission spectrum ex-tends beyond 800 nm, while the former main emission below
Fig. 7. Dose response for the integrated TL signal of LiF:Mg,Cu,P within
400 nm is nearly absent. Generally, it seems that the higher
ranges 250–350 and 350-600 ◦C. Solid lines indicate linear trends.
the dose, the more of the longer-wavelength bands appear.
Unfortunately, for technical reasons, emission spectra weremeasured only up to 350 ◦C, therefore the main emission at
4. Summary
the highest doses was not recorded. Nevertheless, the obtainedresults clearly indicate that at the highest doses different re-
The behaviour of MCP-N (LiF:Mg,Cu,P) detectors at ultra-
combination centres become activated.
high doses has been investigated. While the shape of the
In order to quantify the dose response of the high-temperature
glow-curve at doses ranging from a microgray to a kilogray is
part of the glow-curve, the TL signal was integrated over fixed
practically identical, significant changes are observed at higher
temperature regions: 250–350 and 350–600 ◦C. The depen-
doses. According to these changes in the shape of the glow-
dence of these integrals on dose is presented in The
curve, three sub-ranges of dose can be distinguished: 1–5,
simple (however, not necessarily linear) dose response of the
5–50 and 50–500 kGy. Within the first sub-range, the nominal
new high-temperature peak suggests the possibility of its ap-
main peak 4 at ca. 230 ◦C remains the most prominent one in
plication for dosimetry of ultra-high doses. This would make
the glow-curve, changes being observed only as a growth of
MCP-N a unique detector capable of measuring doses from be-
the high-temperature tail of this main peak. Within the second
low 1 Gy to at least 0.5 MGy, i.e. over a dynamic range of 12
sub-range the glow-curve becomes very complex, consisting
orders of magnitude.
of at least 10 peaks, with a reduced (already saturated) peak 4.
P. Bilski et al. / Radiation Measurements 43 (2008) 315 – 318
Also a new emission spectrum band at 550 nm appears. Within
the third sub-range a completely new peak appears at very hightemperatures. Its position was found to shift towards higher
Bilski, P., 2002. Lithium fluoride: from LiF:Mg,Ti to LiF:Mg,Cu,P. Radiat.
temperatures with dose, in contradiction to standard TL mod-
Prot. Dosim. 100, 199–206.
Bos, A.J.J., 2007. Theory of thermoluminescence. Radiat. Meas. 41,
els. The kinetic energy and frequency factors of this peak also
demonstrate unusual features. These observations make the
Horowitz, Y.S., 1993. LiF:Mg,Ti versus LiF:Mg,Cu,P: the competition heats
new peak very interesting from the theoretical point of view. A
up. Radiat. Prot. Dosim. 47, 135–141.
possible explanation of this phenomenon, based on the model
Mandowski, A., 2007. How to detect trap cluster systems? Radiat. Meas.,
of interacting clusters, was given recently by
Olko, P., Bilski, P., Michalik, V.M., 1994. Microdosimetric analysis of the
At doses around 500 kGy, an emission band beyond
response of LiF thermoluminescent detectors for radiations of different
800 nm emerges in the emission spectrum. This new TL peak
qualities. Radiat. Prot. Dosim. 52, 405–408.
in MCP-N is of considerable theoretical interest and could
Puchalska, M., Bilski, P., 2006. GlowFit: a new tool for thermoluminescence
potentially be exploited in ultra-high-level dosimetry, both
glow-curve deconvolution. Radiat. Meas. 41, 659–664.
subjects being currently pursued.
Velbeck, K.J., Luo, L.Z., Ramlo, M.J., Rotunda, J.E., 2006. The dose–response
of Harshaw TLD-700H. Radiat. Prot. Dosim. 119, 255–258.
This work was partly supported by a research project from
the Polish Ministry of Science over the years 2006–2008(no. N20703931/1541).
Source: http://ajd.czest.pl/~e.mandowska/publikacje/2008rad_mea.pdf
Le coût du hadj sera annoncé la semaine prochaine Lire en page 3 N° 5152 - MERCREDI 22 AVRIL 2015 QUALIFIÉE DE MASCARADEPAR LES PROCHES DU CHEIKH La statue de Benbadis«évacuée» du centre Lire en page 7 Enlèvement de la marchandise prohibée du port sec Anisfer de Rouiba NEUF DOUANIERS SOUS MANDAT DE DÉPÔT Le juge d'instruction de la deuxième chambre près le tribunal de Sidi-M'hamed (Alger), a ordonné hier, la mise en détention préventive de
- Copyright - Il Pensiero Scientifico Editore downloaded by IP 193.205.69.252 Wed, 26 Feb 2014, 17:07:22 LINEE GUIDA Linee guida per la valutazione preoperatoriadel rischio cardiaco e la gestione perioperatoriadel paziente cardiopatico nella chirurgianon cardiaca Task Force per la Valutazione Preoperatoria del Rischio Cardiaco ela Gestione Perioperatoria del Paziente Cardiopatico nella ChirurgiaNon Cardiaca della Società Europea di Cardiologia (ESC) e approvatedalla Società Europea di Anestesiologia (ESA)